首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Development and validation of a gas chromatography/mass spectrometry metabonomic platform for the global profiling of urinary metabolites
Authors:Pasikanti Kishore K  Ho Paul C  Chan Eric C Y
Institution:Department of Pharmacy, Faculty of Science, National University of Singapore, 18 Science Drive 4, Singapore 117543.
Abstract:This paper presents a simple and reliable gas chromatography/mass spectrometry (GC/MS) method for the metabonomic analysis of human urine samples. The sample preparation involved the depletion of excess urea via treatment with urease and subsequent protein precipitation using ice-cold ethanol. An aliquot of the mixture was separated, dried, trimethylsilyl (TMS)-derivatized and 1.0 microL of the derivatized extract was injected into the GC/MS system via split injection (1:10). Approximately 150 putative metabolites belonging to different chemical classes were identified from the pooled human urine samples. All the identified metabolites were selected to evaluate precision and stability of the GC/MS assay. More than 95% of the metabolites demonstrated good reproducibility, with intra-day and inter-day precision values below 15%. Metabolic profiling of 53 healthy male and female urine samples in combination with pattern recognition techniques was performed to further validate the GC/MS metabolite profiling assay. Principal component analysis (PCA) followed by orthogonal partial least squares analysis (OPLS) revealed differences between urinary metabolite profiles of healthy male and female subjects. This validated GC/MS metabolic profiling method may be further applied to the metabonomic screening of urinary biomarkers in clinical studies. Copyright (c) 2008 John Wiley & Sons, Ltd.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号