Determination of mercury in seawater by cold vapour atomic absorption spectrophotometry |
| |
Authors: | Peter Freimann and Diether Schmidt |
| |
Institution: | (1) Labor Sülldorf, Deutsches Hydrographisches Institut, Wüstland 2, D-2000 Hamburg 55, Federal Republic of Germany |
| |
Abstract: | Summary To avoid systematic errors in the sampling and determination of mercury in seawater an integrated sampling-analyzing system has been developed. A 500 ml PTFE-bottle attached to a sampling gear is used as sampling vessel for the cold vapour atomic absorption Spectrophotometry, using nitrogen aeration, SnCl2 reduction, and preenrichment/ purification by amalgamation on finely dispersed gold. The detection limit of this procedure was found to be 0.5 ng·l–1 and the coefficient of variation 4% for 1.5 ng·l–1.
Bestimmung von Quecksilber im Meerwasser durch Kaltdampf-Atomabsorptionsspektralphotometrie Zusammenfassung Zur Vermeidung von systematischen Fehlern bei der Ultraspurenanalyse von Quecksilber im Meerwasser wurde ein integriertes Probenahme-Analysensystem entwickelt. Eine 500 ml PTFE-Flasche dient in Verbindung mit einem Greifarm als Probenahmegefäß für Oberflächenwasser und wird anschließend als Reaktionsgefäß in die Analysenapparatur gesetzt. Die Bestimmung erfolgt mit der Kaltdampf-Atomabsorptionsspektralphotometrie nach Reduktion des Quecksilbers durch SnCl2 und Anreicherung sowie Reinigung mittels Amalgamierung an feinverteiltem Gold. Die Nachweisgrenze dieses Verfahrens beträgt 0,5 ng·l–1, der Variationskoeffizient 4% bei 1,5 ng·l–1. |
| |
Keywords: | |