Reassessment of doses in TLD100 by measuring the residual dose |
| |
Authors: | U German A Abraham M Weinstein Z B Alfassi |
| |
Institution: | (1) Nuclear Research Center Negev, P.O. Box 9001, 84190 Beer Sheva, Israel;(2) Ben-Gurion University of the Negev, 84105 Beer Sheva, Israel |
| |
Abstract: | Second readouts of TLD 100 crystals exposed to radiation are used to reassess high doses. When the second readout is performed
by a regular readout cycle of 13.3 s, the accuracy of the estimations is too low to be practical for low doses in the 10–100
mSv range. A simple and efficient method was studied, which consists of expanding the heating time to 30 s. The efficiency
relative to a 13.3 readout time is about three timesbetter, and also the background is lower, thus the reassessment efficiency
is significantly improved. |
| |
Keywords: | |