首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Reassessment of doses in TLD100 by measuring the residual dose
Authors:U German  A Abraham  M Weinstein  Z B Alfassi
Institution:(1) Nuclear Research Center Negev, P.O. Box 9001, 84190 Beer Sheva, Israel;(2) Ben-Gurion University of the Negev, 84105 Beer Sheva, Israel
Abstract:Second readouts of TLD 100 crystals exposed to radiation are used to reassess high doses. When the second readout is performed by a regular readout cycle of 13.3 s, the accuracy of the estimations is too low to be practical for low doses in the 10–100 mSv range. A simple and efficient method was studied, which consists of expanding the heating time to 30 s. The efficiency relative to a 13.3 readout time is about three timesbetter, and also the background is lower, thus the reassessment efficiency is significantly improved.
Keywords:
本文献已被 SpringerLink 等数据库收录!