首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Carbon as Reference for Normalizing Infrared Photoacoustic Spectra
Authors:M J D Low  G A Parodi
Institution:Department of Chemistry New York University 4 Washington Place New York , N. Y, 10003
Abstract:Photoacoustic spectra of solids1 are usually “source compensated,” i. e., normalized for differences in source emission at different wavelengths, and such compensation is done by comparing the photoacoustic spectrum of a sample to the emission spectrum of the source measured directly with, say, a photocell,2 power meter,3 or photodiode,4 or to the photoacoustic spectrum of “carbon” recorded under the same conditions as that of the sample. A variety of carbons has been used for this purpose, including carbon black on silica,5 soot from a gas torch,6 and acatylene-sooted bakelite,7 such a reference substance being assumed to be a flat black absorber. That assumption appears to be valid, as indicated by the close
Keywords:Photoacoustic Spectroscopy  Infrared  Carbon  Surface Species
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号