首页 | 本学科首页   官方微博 | 高级检索  
     


Determination of Se, As, Cu, Pb, Cd, Zn and Mn by anodic and cathodic stripping voltammetry in marine environmental matrices in the presence of reciprocal interference. Proposal of a new analytical procedure
Authors:Clinio Locatelli  Giancarlo Torsi
Affiliation:Department of Chemistry ‘G. Ciamician’, University of Bologna, Via Selmi 2, I-40126 Bologna, Italy
Abstract:Arsenic(III), selenium(IV), copper(II), lead(II), cadmium(II), zinc(II) and manganese(II) have been determined in environmental matrices by differential pulse cathodic (DPCSV) and anodic (DPASV) stripping voltammetry. The voltammetric measurements were carried out using a conventional three-electrode cell and the ammonia–ammonium chloride buffer (pH 9.4) as the supporting electrolyte. The analytical procedure was verified by the analysis of the standard reference materials: Sea Water BCR-CRM 403; Lagarosiphon Major BCR-CRM 060; and Cod Muscle BCR-CRM 422. The precision, expressed as relative standard deviation, and the accuracy, expressed as relative error, were, in all cases, lower than 5%; the detection limits, for each element in the experimental conditions employed, was approximately 10−9 M. The standard addition technique significantly improved the resolution of the voltammetric method, even in the case of very high metal concentration ratios.
Keywords:Stripping voltammetry   Metal interference   Marine environmental matrices
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏