首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Glucose enzyme electrode with extended linearity: Application to undiluted blood measurements
Authors:William H Mullen  Fiona H Keedy  Stephen J Churchouse  Pankaj M Vadgama
Institution:Department of Physical Chemistry, The University, Newcastle upon Tyne NE1 7RU Great Britain;Department of Clinical Biochemistry and Metabolic Medicine, The Medical School, Framlington Place, Newcastle upon Tyne, NE2 4HH Great Britain
Abstract:An enzyme electrodes is described for glucose determination in unstirred, undiluted whole blood. The system comprises an H2O2-detecting electrode upon which is placed a membrane laminate incorporating glucose oxidase. The external membrane was pretreated with methyltrichlorosilane. The electrode response was linearly dependent on glucose concentration up to 50 mmol l?1 glucose, it had a decreased dependence on dissolved oxygen concentrations and gave response times of 30–90 s. Whole blood glucose measurements correlated well with a routine spectrophotometric method.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏