Pulsed coulometric detection of carbohydrates at a constant detection potential at gold electrodes in alkaline media
Authors:Glen G Neuburger  Dennis C Johnson
Affiliation:Department of Chemistry, Iowa State University, Ames, IA 50011 U.S.A.
Abstract:A significant increase in the signal-to-noise ratio for the pulsed amperometric detection (PAD) of carbohydrates at gold electrodes is obtained by increasing the length of the current integration period (ti) from the traditional value of 16.7 ms (i.e., 160 Hz). For ti > 16.7 ms, the integrated response (q, coulombs) is plotted as the signal. This pulsed coulometric detection (PCD) is applied in a flow-injection system. For ti = 500 ms, the detection limit with the instrumentation used is 1 μM (S/N = 2) for glucose which is a significant improvement on the value 35 μM found with PAD. The absolute detection limits for glucose and sucrose are ca. 50 pmol and 125 pmol, respectively, in 50-μl samples. Calibration plots (qp vs. Cb) for PCD are linear over significantly larger dynamic ranges than those observed for PAD because of the lower detection limits.