Pneumatopotentiometric determination of nanogram amounts of cyanide
Authors:František Opekar
Affiliation:The UNESCO Laboratory of Environmental Electrochemistry, J. Heyrovský Institute of Physical Chemistry and Electrochemistry, Czechoslovak Academy of Sciences, Jilská 16, 110 00 Prague 1 Czechoslovakia
Abstract:Hydrogen cyanide is liberated from aqueous samples by reaction with sulphuric acid and transferred by a stream of nitrogen to a silver porous membrane electrode. Some HCN passes through the membrane into an alkaline dicyanoargentate solution; the cyanide ion produced causes a decrease in the equilibrium Ag+ concentration and the change of potential is related to the amount of cyanide in the sample. The detection limit is 3.0 ng ml?1 cyanide in the injected solution; the relative standard deviation is 0.82% for 17 ng of cyanide. Sulphide interferes (as H2S) but can be removed on a lead acetate column.