首页 | 本学科首页   官方微博 | 高级检索  
     检索      


An automated spectrophotometric field monitor for water quality parameters : Determination of nitrate
Authors:JRichard Clinch  Paul J Worsfold  Harry Casey
Institution:Department of Chemistry, University of Hull, Hull HU6 7RX Great Britain;Freshwater Biological Association, River Laboratory, East Stoke, Wareham, Dorset BH20 6BB Great Britain
Abstract:An automated spectrophotometric field monitor is described for the determination of nitrate in river water. The bias, precision and linear calibration range of the monitor were determined in the laboratory. Samples taken from six locations on the River Frome in Dorset were analyzed for nitrate content by the monitor, and the results obtained were in good agreement with those obtained by two spectrophotometric reference methods. The relative standard deviation for 672 replicate injections of a 7.7 mg l?1 NO3 standard over a 14-day period was 1.4%, during which time 2.51 of the colour reagent and 2.5 1 of the ammonium chloride solution were consumed. The linear calibration range of the procedure was 0–12 mg l?1 NO3-N. Results are also presented for a four-day field trial on the River Frome.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏