Comparative study of potentiometric and amperometric tissue-based electrodes for oxalate |
| |
Authors: | Tekum Fonong |
| |
Institution: | Department of Chemistry, Illinois State University, Normal, IL 61761 U.S.A. |
| |
Abstract: | Acetone-precipitated pulp from banana skins is physicall entrapped at the tip of a carbon dioxide gas-sensor and on a hydrogen peroxide sensor probe to determine oxalate potentiometrically and amperometrically in aqueous solution and inurine. The enzyme present in the tissue is oxalate oxidase. The potentiometric response has a slope of 47–50 mV/decade for 1 × 10?4 M–2 × 10?3 M oxalate with a detection limit of 2 × 10?5 M. The amperometric response is linear for 2 × 10t-5–3 × 10?4 M oxalate with a dectection limit of 2 × 10?6 M. Average recoveries of oxalate added to aqueous samples were 96.2% and 98.0%, and average relative standrd deviations were 3.8% and 3.6% for the potentiometric and amperometric systems, respectively. Oxalate was determined in six control urine samples, with relative errors of about 2.5%, by both electrode systems after a simple clean-up. |
| |