首页 | 本学科首页   官方微博 | 高级检索  
     


Mediated amperometric biosensors for d-galactose,glycolate and l-amino acids based on a ferrocene-modified carbon paste electrode
Authors:Jonathan M. Dicks  William J. Aston  Graham Davis  Anthony P.F. Turner
Affiliation:Bioelectronics Division, Biotechnology Centre, Cranfield Institute of Technology, Cranfield, Bedfordshire, MK43 0AL Great Britain
Abstract:Direct-current cyclic voltammetry is used to investigate the suitability of a ferrocene derivative as a mediator with galactose, glycolate and l-amino acid oxidases. The three enzymes coupled catalytically to ferrocene monocarboxylic acid exhibiting homogeneous second-order rate constants in the range 0.4 × 105 to 8.5 × 105 l mol?1 s?1. Enzyme electrodes which responded to d-galactose, glycolate or l-amino acids were constructed. The appropriate oxidase was retained behind a dialysis membrane at a carbon paste electrode containing the poorly soluble derivative 1,1′-dimethylferrocene. All the electrodes responded rapidly to millimolar concentrations of their respective substrates producing 95% of the steady-state current response in <2 min. This general method of biosensor construction should be widely applicable to oxidases and other oxidoreductase enzymes.
Keywords:
本文献已被 ScienceDirect 等数据库收录!