Abstract:
A capillary scale reverse phase liquid chromatography (LC) system using a super hot water eluent is described. The system, constructed in-house from readily available components, has been shown to operate at temperatures as high as 370 °C and pressures in excess of 10 000 psi. The capability of the system is demonstrated with the separation of a mixture of polar and non-polar benzene derivatives on polybutadiene and elemental carbon modified zirconia packings with or without temperature gradients. Six benzene derivatives can be separated in 2 min.