Determination of iron(II) and iron(III) by flow injection and amperometric detection with a glassy carbon electrode |
| |
Authors: | J.W. Dieker W.E. Van Der Linden |
| |
Affiliation: | Laboratory for Analytical Chemistry, University of Amsterdam, Niiewe Achtergracht 166,1018 WV AmsterdamThe Netherlands |
| |
Abstract: | Flow injection analysis can be used for the determination of both iron(II) and iron(III) with an amperometric detector. The flow-through cell contains a glassy carbon electrode. Selection of the appropriate voltammetric technique, choice of the indication potentials, sample size, composition of the carrier stream, etc., are discussed. The limit of determination is about 10-6 M; the calibration curves are linear in the concentration ranges 10-3–10-5 M for iron(III) and 5 × 10-4–10-5 M for iron(II). To illustrate the potentialities of the proposed method, standard rocks have been analysed. |
| |