首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Determination of iron(II) and iron(III) by flow injection and amperometric detection with a glassy carbon electrode
Authors:JW Dieker  WE Van Der Linden
Institution:Laboratory for Analytical Chemistry, University of Amsterdam, Niiewe Achtergracht 166,1018 WV AmsterdamThe Netherlands
Abstract:Flow injection analysis can be used for the determination of both iron(II) and iron(III) with an amperometric detector. The flow-through cell contains a glassy carbon electrode. Selection of the appropriate voltammetric technique, choice of the indication potentials, sample size, composition of the carrier stream, etc., are discussed. The limit of determination is about 10-6 M; the calibration curves are linear in the concentration ranges 10-3–10-5 M for iron(III) and 5 × 10-4–10-5 M for iron(II). To illustrate the potentialities of the proposed method, standard rocks have been analysed.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号