Simultaneous determination of nitrate and nitrite by flow injection analysis |
| |
Authors: | M.F. Giné H. Bergamin F E.A.G. Zagatto B.F. Reis |
| |
Affiliation: | Centro de Energia Nuclear na Agricultura, CEP, 13.400 Piracicaba S. PauloBrasil |
| |
Abstract: | An automatic method for the simultaneous determination of nitrate and nitrite by flow injection analysis is described. Nitrate is reduced to nitrite with a copperized cadmium column. Nitrite is diazotized and coupled with N-(l-naphthyl)ethylenediammonium dichloride. The merging zones approach is used to minimize reagent consumption. The injector system is arranged so that two peaks are obtained, one corresponding to nitrite and the other to nitrite plus nitrate. A sampling rate of about 90 samples per hour is possible; the precision is better than 0.5% for nitrite in the range 0.1–0.5 mg lt and 1.5% for nitrate in the range 1.0–5.0 mg lt |
| |
Keywords: | |
本文献已被 ScienceDirect 等数据库收录! |