首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Calibration of a Gardon Sensor in a High-Temperature High Heat Flux Stagnation Facility
Authors:P Stathopoulos  F Hofmann  T Rothenfluh  Ph R von Rohr
Institution:1. Institute of Process Engineering, Swiss Federal Institute of Technology (ETH) , Zurich , Switzerland;2. PanGas AG Industriepark 10 , Dagmersellen , Switzerland
Abstract:A calibration methodology developed for custom-made heat flux sensors used for the investigation of hydrothermal spallation drilling is presented. An air jet stagnation convection calibration system allowing independent control of jet velocity and temperature, cooling water flow and temperature, stand off distance, and nozzle diameter was built. A thermopile device was used as reference heat flux sensor. Air jets with velocities up to 400 m/s, exit temperatures of 760°C, and heat fluxes up to 0.6 MW/m2 can be reached. The standard deviation of a reference heat flux measurement is below 2%, and overall uncertainty achieved is less than 5.5%.
Keywords:calibration system  heat flux sensor  Gardon gauge  free-stream temperature  spallation drilling
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号