首页 | 本学科首页   官方微博 | 高级检索  
     


Flow injection determination of peroxide value in edible oils using triiodide detector
Authors:Bahruddin Saad  Wan Tatt Wai  Boey Peng Lim  Muhammad Idiris Saleh
Affiliation:School of Chemical Sciences, Universiti Sains Malaysia, 11800 Penang, Malaysia
Abstract:A flow injection analysis (FIA) method for the determination of peroxide value (PV) in edible oils is described. Oil sample (undiluted) and KI reagent were aspirated into a homemade reaction chamber where the redox reaction between iodide in the aqueous phase and hydroperoxides in the oil was effected by applying a short (typically 30 s) vortex action. After allowing for the emulsified oil phase to be separated from the aqueous phase (bottom layer), an aliquot of the aqueous phase containing triiodide was next aspirated to the surface of a triiodide-selective membrane for detection. The optimized FIA procedure is linear over 0.35-28.0 PV (mequiv. O2/kg) with a detection limit of 0.32 PV. Exhibiting good reproducibility (R.S.D. of 2.7% (n = 8) for the determination of 1.1 PV) and sampling rate of 80 samples h−1, the proposed method, unlike previous FIA procedures, completely eliminated the use of organic solvents (except the use of 2-propanol for cleaning of reaction chamber). Excellent correlation (R2 = 0.9949) between the proposed method and the manual official AOCS method was found when applied to the determination of PV in diverse type of edible oils (n = 20).
Keywords:Peroxide value   Edible oil   Flow injection analysis   Triiodide electrode
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号