Organotin compounds: an ionophore system for fluoride ion recognition |
| |
Authors: | Chandra Sudeshna Růzicka Ales Svec Petr Lang Heinrich |
| |
Affiliation: | Technische Universit?t Chemnitz, Fakult?t für Naturwissenschaften, Institut für Chemie, Lehrstuhl Anorganische Chemie, Strasse der Nationen 62, 09111 Chemnitz, Germany. |
| |
Abstract: | Ion-selective properties were established for membrane electrodes prepared by using organotin compounds of type (LCNRSnF2)n, (R = n-Bu (I), = Ph (II)) and (LCNSnF3)n (III) (LCN = C6H4(CH2NMe2)-2). Electrodes formulated with the optimized membranes containing the organotin compounds I-III as ionophores and sodium tetraphenylborate (10-30%) exhibited high selectivity for fluoride over other anions. An electrode prepared with ionophore II using dibutyl phthalate as the plasticizer and 15% sodium tetraphenylborate (NaTPB) as anion additive, possesses the best potentiometric response characteristics. It shows a detection limit of 7.9 × 10−7 M with a slope of 62.7 mV decade−1 of activity in buffer solutions of pH 5.5. The interference from other anions is suppressed under this optimized measurement conditions. An entirely non-Hofmeister selectivity sequence (F− > CH3COO− > Cl− > I− ∼ Br− >ClO4− > NO2− > NO3− > SCN−) with remarkable preference towards fluoride is obtained. The influence on the electrode performances by anion additive was studied, and the possible response mechanism was investigated by UV-vis spectra. The electrode has been used for direct determination of fluoride in drinking mineral water with satisfactory results. |
| |
Keywords: | Organotin compounds Fluoride selective electrode Ionophores Anion recognition |
本文献已被 ScienceDirect PubMed 等数据库收录! |
|