An automated high precision calorimeter for the temperature range 200K–400K. Part 1: Design and performance
Authors:I S Williams  R Street  E S R Gopal
Affiliation:(1) Research School of Physical Sciences, Australian National University, 2600 Canberra, A C T, Australia;(2) Present address: University of Western, Australia;(3) Present address: Physics Department, Indian Institute of Science, 560 012 Bangalore
Abstract:A fully automated calorimeter has been designed and tested over the temperature range 200K–400K. The system may be used for measurements with an absolute accuracy of 0.2% of samples of approximate mass 50g and thermal capacity 15 JK?1. The temperature of the sample is determined by a quartz crystal thermometer of resolution 100 µK which is not in direct thermal contact with the sample. The performance of the system is illustrated by results obtained on high purity copper, distilled water and K2PbCu(NO2)6 which exhibits very sharp first order phase transitions at about 273.4K and 281.8K.
Keywords:Specific heat  calorimeter  phase transitions