首页 | 本学科首页   官方微博 | 高级检索  
     


Atmospheric pressure chemical ionization mass spectrometry of aldehydes in biological matrices
Authors:Nagy Kornél  Pollreisz Ferenc  Takáts Zoltán  Vékey Károly
Affiliation:Institute of Chemistry, Chemical Research Center, Hungarian Academy of Sciences, H-1025 Pusztaszeri út 59-67, Budapest, Hungary.
Abstract:Application of high-performance liquid chromatography (HPLC) and tandem mass spectrometry (MS/MS) for detection of aldehydes in biological samples such as blood, yoghurt, and milk, is reported. Sample preparation is easy, and the presented method is both sensitive and selective. It is based on the widely used dinitrophenylhydrazine derivatization, followed by extraction with n-hexane and a simple reversed-phase HPLC separation. Detection is performed by atmospheric pressure chemical ionization (APCI) in negative ion mode, with detection limits in the low picogram range. Using MS/MS, acetone and propionaldehyde can clearly be distinguished, facilitating propionaldehyde quantitation even in the presence of high acetone levels. Quantitation by direct MS/MS is also feasible, well suited for high-throughput applications, although not as accurate as using HPLC/MS/MS.
Keywords:
本文献已被 PubMed 等数据库收录!