首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Use of Interrupted Helium Flow in the Analysis of Vapor Samples with Flowing Atmospheric-Pressure Afterglow-Mass Spectrometry
Authors:Andrew P Storey  Offer M Zeiri  Steven J Ray  Gary M Hieftje
Institution:1.Department of Chemistry,Indiana University,Bloomington,USA;2.Nuclear Research Center Negev,Beer-Sheva,Israel;3.Department of Chemistry,University at Buffalo, The State University of New York,Buffalo,USA
Abstract:The flowing atmospheric-pressure afterglow (FAPA) source was used for the mass-spectrometric analysis of vapor samples introduced between the source and mass spectrometer inlet. Through interrupted operation of the plasma-supporting helium flow, helium consumption is greatly reduced and dynamic gas behavior occurs that was characterized by schlieren imaging. Moreover, mass spectra acquired immediately after the onset of helium flow exhibit a signal spike before declining and ultimately reaching a steady level. This initial signal appears to be due to greater interaction of sample vapor with the afterglow of the source when helium flow resumes. In part, the initial spike in signal can be attributed to a pooling of analyte vapor in the absence of helium flow from the source. Time-resolved schlieren imaging of the helium flow during on and off cycles provided insight into gas-flow patterns between the FAPA source and the MS inlet that were correlated with mass-spectral data.
Graphical Abstract ?
Keywords:
本文献已被 SpringerLink 等数据库收录!