首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Thiocyanate-selective electrode based on unsymmetrical benzoN(4) nickel(II) macrocyclic complexes
Authors:Abbaspour A  Kamyabi M A  Esmaeilbeig A R  Kia R
Institution:Chemistry Department, College of Science, Shiraz University, Shiraz 71454, Iran.
Abstract:Unsymmetrical benzoN(4) nickel(II) macrocyclic complexes have been explored as an ionophore for preparing PVC-based membrane electrodes selective to SCN(-). The resulting electrode exhibited the best working concentration range (1.4x10(-7)-0.1 M) with a Nernstian slope of (-59.7+/-0.9 mV per decade) in the wide pH range of 3.5-10.5. The proposed electrode has a very fast response time (5 s) and high selectivity for thiocyanate over variety of anions. This sensor was successfully applied as an indicator electrode to potentiometric titration of thiocyanate ion with AgNO(3) and for direct determination of the amount of thiocyanate in urine and saliva samples in order to distinguish between smokers and nonsmokers.
Keywords:
本文献已被 PubMed 等数据库收录!