首页 | 本学科首页   官方微博 | 高级检索  
     


Development and validation of a differential scanning calorimetry purity determination method for polycyclic aromatic hydrocarbons
Authors:Vikram Kestens  Gert Roebben  Thomas Linsinger
Affiliation:(1) Institute for Reference Materials and Measurements (IRMM), Joint Research Centre, European Commission, 2440 Geel, Belgium
Abstract:Based on a standard test method for purity by differential scanning calorimetry (DSC), ASTM E 928, a purity determination method for highly pure polycyclic aromatic hydrocarbons (PAHs) has been developed and validated. The robustness of the developed method was investigated by determining, under varying measurement conditions, the purity of two PAH certified reference materials (CRMs), benzo[c]phenanthrene and dibenzo[a,h]anthracene. The repeatability and intermediate precision of the developed method was determined by analysing the purity of benzo[c]phenanthrene and dibenzo[a,h]anthracene and PAH candidate CRMs indeno[1,2,3-c,d]pyrene, 6-methylchrysene and benzo[a]pyrene. The trueness of the method was studied using the same (candidate) CRMs and a series of 42 other PAH CRMs. For each of the five (candidate) CRMs, a full measurement uncertainty budget was developed. Also for PAH materials for which the DSC purity determination method has not been explicitly validated, the relative expanded measurement uncertainty was estimated.