A high resolution 4π alpha spectrometer with silicon surface barrier detectors
Authors:J E Noakes  J L Duggan
Institution:(1) Geochronology Laboratory, University of Georgia, 30602 Athens, Georgia, (USA);(2) Physics Department, North Texas State University, 76203 Denton, Texas, (USA)
Abstract:An alpha counting system has been designed which gives counting geometry of nearly 4π with resolution capabilities of 25 keV or better. The alpha source is electro-deposited onto a thin gold foil which is sandwiched between two silicon surface barrier detectors. Alpha pulses from the two detectors are amplified, normalized and summed as a single pulse height spectrum. The techniques for the electrodeposition of the alpha activity onto the thin gold foil are described.