(1) Department of Chemistry, Faculty of Science, Bu-Ali Sina University, Hamadan, Iran
Abstract:
A method for the determination of sulfide based on its interference with the determination of Hg by cold vapor atomic absorption spectrometry is described. The decrease in mercury absorbance at 253.7nm is proportional to the concentration of sulfide over the range of 10–320ngmL–1. The limit of detection was found to be 7ngmL–1 and the relative standard deviation (R.S.D.) for the determination of different concentrations of sulfide was in the range of 1.8–2.2%. This method was applied to the determination of sulfide in whole human blood after gas-phase separation.