Improvement of the Detection Limit in Capillary Gas Chromatographic Trace Analysis of C2–C4 Hydrocarbons by Post Column Cryogenic Trapping
Authors:Maurice Habram  Thomas Welsch
Abstract:A method for the improvement of signal-to-noise ratios and hence detection limits in capillary gas chromatographic trace analysis of C2–C4 hydrocarbons was developed. It is based on the post column cryogenic trapping of the separated hydrocarbons using liquid nitrogen as coolant and its fast reinjection into the detector by rapid heating of the capillary tubing used for trapping. The improved signal-to-noise ratios were applied to decrease the sample volume and/or to lower the detection limit 10 to 27 times. The concentrations of these hydrocarbons in an urban air sample determined without and with post column cryogenic trapping were in good agreement but the precision was better when applying the cryogenic trapping mode.
Keywords:Capillary GC trace analysis of C2–  C4 hydrocarbons  signal-to-noise ratio, improvement  post column cryogenic trapping  thermal remobilization