Use of blotting-paper collectors for the separation and X-ray fluorescent determination of traces of heavy metals (Cu, Hg, Ag) in water and waste water
Authors:Zbigniew Gregorowicz  Henryk Stec and Jerzy Ciba
Institution:(1) Inst. of Anal. and General Chemistry, Silesian Techn. University, PL 44-101 Gliwice, Poland
Abstract:Summary Traces of Hg, Ag and Cu were separated in the range of 0.5–500 Mgrg with blotting-paper collectors and were subsequently determined directly with energy dispersive X-ray fluorescence. For comparison, spectrophotometric determinations were carried out as well. The results were in good agreement. The collectors contained 0.2 mM ZnS, in case of Ag a thioureide collector was also employed. Limits of detection were 0.2 Mgrg Hg, 0.4 Mgrg Cu and 0.15 Mgrg Ag for the ZnS collector and 0.13 Mgrg Ag for the thioureide collector.
Verwendung von Flie\papier-Kollektoren zur Abtrennung und röntgenfluorescenzspektrometrischen Bestimmung von Schwermetallspuren (Cu, Hg, Ag) in Wasser und Abwasser
Zusammenfassung Hg, Ag und Cu im Bereich von 0,5–500 Mgrg wurden auf Flie\papier-Kollektoren abgetrennt und direkt durch energiedispersive Röntgenfluorescenzanalyse bestimmt. Die Ergebnisse wurden durch Spektralphotometrie überprüft. Die Kollektoren enthielten 0,2 mM ZnS, für Ag statt dessen auch Thioureid. Alle Ergebnisse waren in guter übereinstimmung. Die Nachweisgrenzen für den ZnS-Kollektor betrugen 0,2 Mgrg Hg, 0,4 Mgrg Cu, 0,15 Mgrg Ag; für den Thioureid-Kollektor 0,13 Mgrg Ag.
Keywords:Best  von Silber  Quecksilber  Kupfer in Wasser  Abwasser    ntgenfluorescenz-Spektrometrie/Spektralphotometrie  Flie\papier-Kollektor zur Abtrennung (ZnS  Thioureid)