首页 | 本学科首页   官方微博 | 高级检索  
     


Achromatic customized intraocular lens based on an individual pseudophakic eye model
Authors:Feng Rao  Zhao-Qi Wang  Yan Wang
Affiliation:a Institute of Modern Optics, Nankai University, Tianjin 300071, PR China
b Key laboratory of Opto-electronic Information Science and Technology, Ministry of Education, PR China
c Tianjin Eye Hospital, Refractive Surgery Center, Tianjin 300020, PR China
Abstract:Customized hybrid diffractive-refractive intraocular lens (IOL) has been designed based on an individual pseudophakic eye model constructed with the measured corneal topography data and the eye's axial lengths. The detailed designing and optimizing procedure is given. It is found that an eye with achromatic customized IOL would achieve super modulation transfer function (MTF) in both the scotopic and photopic conditions. The variation of MTF as colors is negligible, the chromatic focus shift is no more than 0.1 D, and the depth of focus is 3.0 D with quite high MTF. The achromatic customized IOL provides excellent visual performance for cataract patients. Moreover, individual pseudophakic eye model can be an effective tool for evaluating the optical performance of pseudophakic eye and the influence of clinical surgery deviation.
Keywords:Visual optics   Cataract surgery   IOL   Individual eye model   MTF