首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Determination of isotropic layer parameters from spatiotemporal signals of an ultrasonic array
Authors:S A Titov  R G Maev
Institution:1. Emanuel Institute of Biochemical Physics, Russian Academy of Sciences, Moscow, 119991, Russia
Abstract:The paper discusses a method for measuring the velocities and attenuation of longitudinal and transverse ultrasonic waves and the density and thickness of the isotropic layer with an array placed in an immersion liquid parallel to the sample. The method is based on the recording of the total spatiotemporal signal of the array and its expansion into a spatial spectrum of pulse plane wave response. The ultrasonic velocity and sample thickness depend on the response delay of the plane wave in the layer from the transverse projection of the slowness vector. The density and attenuation are determined from the behavior of the amplitudes of spectral responses. To confirm this method in experiment, the parameters of a polystyrene plate have been measured using a linear 32-element array with a central frequency of 17 MHz.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏