首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Frequency-stabilized cavity ring-down spectrometer for high-sensitivity measurements of water vapor concentration
Authors:JT Hodges  D Lisak
Institution:(1) Chemical Science and Technology Laboratory, National Institute of Standards and Technology, 100 Bureau Drive, Gaithersburg, MD 20899, USA;(2) Instytut Fizyki, Uniwersytet Mikołaja Kopernika, ul. Grudziadzka 5/7, 87-100 Toruń, Poland
Abstract:We present a portable spectrometer that uses the frequency-stabilized cavity ring-down spectroscopy technique capable of high-precision measurements of trace water vapor concentration. Measuring one of the strongest rovibrational transitions in the ν13 water vapor combination band near ˜ν=7181.156 cm-1, we compare spectroscopic and thermodynamic determinations of trace water vapor in N2, and find systematic differences attributable to water vapor background effects and/or uncertainties in line intensities. We also compare the frequency-stabilized ring-down method with other cavity ring-down approaches that are based on unstabilized probe lasers and unstabilized ring-down cavities. We show that for the determination of water vapor concentration, the frequency-stabilized cavity ring-down method has the minimum measurement uncertainty of these techniques. The minimum noise-equivalent absorption coefficient of the spectrometer was 1.2×10-10 cm-1 Hz-1/2, which further corresponds to a minimum detectable water vapor mole fraction equal to 0.7×10-9 for an absorption spectrum of 10 minutes duration. PACS 33.20.-t; 33.70.Jg; 33.70.Fd; 42.62.Fi
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏