Systematic errors in the measurement of heat by flow microcalorimetry |
| |
Authors: | Veronica M Poore Anthony E Beezer |
| |
Institution: | Department of Chemistry, Chelsea College, University of London, London SW3 6LX Gt. Britain |
| |
Abstract: | It is shown that a number of systematic errors must be considered when performing heat measurements by flow microcalorimetry because the nature of the flow technique is such that substantial heat loss can be incurred. The conventional procedure of electrical calibration is found to be an inadequate correction parameter. Equations to account for the effects of thermal disequilibrium are derived from the basic principles incorporated in the Tian equation. The predicted relationships are tested by simple experiments and shown to be correct. The various correction parameters are measured for a wide range of flow rate and heat input conditions. A composite equation is presented which allows for the correction of heat loss while deconvoluting electrical heat from a heat of reaction. The total heat output rate from a flow calorimeter can be calculated for most experimental conditions by reference to this equation and to the tabulated correction values. |
| |
Keywords: | To whom correspondence should be addressed |
本文献已被 ScienceDirect 等数据库收录! |