In-house validation of liquid chromatography tandem mass spectrometry for determination of semicarbazide in eggs and stability of analyte in matrix |
| |
Authors: | K. Stastny K. Frgalova A. Hera M. Vass M. Franek |
| |
Affiliation: | 1. Institute for the State Control of Veterinary Biologicals and Medicaments, Hudcova 56a, 621 00 Brno, Czech Republic;2. Veterinary Research Institute, Hudcova 70, 621 00 Brno, Czech Republic |
| |
Abstract: | An improved LC-MS/MS method for the determination of semicarbazide in whole egg is described. Waters OASIS-MCX cation exchange purification cartridges increased the sensitivity for analysis by LC-MS/MS. The validation study was carried out according to criteria and requirements of Commission Decision 2002/657/EC for confirmatory analysis and provided the data as follows: The correlation coefficient for the matrix calibration curve, in the range of 0–5 μg kg−1, was r = 0.9968. The detection capability and decision limit, measured according to ISO11843-2, were CCα = 0.20 μg kg−1 and CCβ = 0.25 μg kg−1. Repeatability (CVSr) and within-laboratory reproducibility (CVSwr) determined for the concentration levels of 0.2, 0.5 and 1.0 μg kg−1 SEM ranged from 11.9 to 5.7% and 11.8 to 6.3%, respectively. The validated method was applied to investigate SEM stability in incurred materials (egg homogenates) during long-term storage at −20 °C and 4 °C. The study proved by a two-sampling test that SEM at levels of 17. 7, 1.2, 10.6 and 0.47 μg kg−1 was stable for up to 12 months. |
| |
Keywords: | Semicarbazide Egg LC-MS/MS validation Incurred material Stability study |