首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Determination of true selectivity coefficients of neutral carrier calcium selective electrode
Authors:Tomasz Sokalski  Magdalena Maj-Żurawska  Adam Hulanicki
Institution:(1) Department of Chemistry, Warsaw University, PL-02-093 Warsaw, Poland
Abstract:Literature data concerning selectivity coefficients of neutral carrier calcium selective electrode show large discrepancies up to several orders of magnitude for sensors containing the same active component. It can be shown theoretically that determination of selectivity coefficients in non-complexing media gives too high values because of additional sources of the main ion in the membrane vicinity. Determination of selectivity coefficients in solutions containing a complexing agent gives the possibility to evaluate true selectivity coefficients not influenced by additional factors. Measurements performed in complexing as well as in non-complexing media and calculation of the detection limit indicate a good agreement between expectations and experimental results. This has been confirmed using the calcium selective electrode with the neutral carrier ETH 1001 witho-nitrophenyl octyl ether or dioctylsebacate as mediators. As interferents sodium, potassium and magnesium ions were tested.
Keywords:calcium selective electrode  selectivity coefficients
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏