首页 | 本学科首页   官方微博 | 高级检索  
     


A computer controlled continuous dilution dilatometer for the determination of solute molar volumes and direct measurement of reaction volumes in solution
Authors:Simon Johnstone  John I. Sachinidis  Richard D. Shalders  Peter A. Tregloan
Affiliation:(1) School of Chemistry, The University of Melbourne, 3052 Parkville, Australia
Abstract:The design and operation of a computer controlled continuous dilution dilatometer utilizing a diode array detection system is presented. To test the accuracy and reproducibility of the technique, apparent molar volumes of dilute aqueous solutions of potassium ferricyanide and potassium ferrocyanide have been measured. To provide supporting data for high pressure studies of reaction volumes and molar volume change in solutions of high ionic strength, study of these electrolytes was extended by dilution into 0.1M and 1.0M potassium nitrate solutions rather than extrapolation to zero ionic strength as is normal. In addition, the system has been used to determine reaction volumes of some acid-base reactions by direct measurement, using the equipment as an automatic titration dilatometer.
Keywords:Molar volume  reaction volume  dilatometry  diode array detection  K3Fe(CN)6  K4Fe(CN)6  HN3
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏