Abstract:Scanning calorimetric methods permit determination of heat capacities at high temperatures up to 1600°C. For disk systems with power compensation application limits are in order of 700°C, and for cylindrical systems with electrical calibration up to 1000°C. For the high temperature range above 1000°C DSC plates and a cylindrical calorimetric systems based on the CALVET principle ('MULTI HTC’) are known. For cylindrical calorimetric systems the precision of the Cp data is between 2 and 5% even at high temperatures without any requirements on the kind and shape of samples. These results are better than data provided by DSC plate systems. This revised version was published online in July 2006 with corrections to the Cover Date.
Keywords:DSC  heat capacity  high temperature