Abstract: | Scanning calorimetric methods permit determination of heat capacities at high temperatures up to 1600°C. For disk systems
with power compensation application limits are in order of 700°C, and for cylindrical systems with electrical calibration
up to 1000°C. For the high temperature range above 1000°C DSC plates and a cylindrical calorimetric systems based on the CALVET
principle ('MULTI HTC’) are known. For cylindrical calorimetric systems the precision of the Cp data is between 2 and 5% even
at high temperatures without any requirements on the kind and shape of samples. These results are better than data provided
by DSC plate systems.