红外光学透镜焦距测量 |
| |
引用本文: | 陈磊,高志山,何勇. 红外光学透镜焦距测量[J]. 光子学报, 2004, 33(8): 986-988 |
| |
作者姓名: | 陈磊 高志山 何勇 |
| |
作者单位: | 南京理工大学电子工程与光电技术学院,南京,210094;南京理工大学电子工程与光电技术学院,南京,210094;南京理工大学电子工程与光电技术学院,南京,210094 |
| |
摘 要: | 采用CO2激光器作为光源,热释电摄像机作为探测器,采集了单缝衍射图像.根据单缝衍射原理,测量了不同焦距的一组红外光学锗透镜的单色焦距,给出了实测结果.讨论了影响测试的主要误差因素.通过计算采样数据的调制传递函数,精确确定被测透镜焦平面的位置.介绍了图像采集系统长度尺寸的精确标定方法.
|
关 键 词: | 红外 锗透镜 焦距 |
收稿时间: | 2003-07-16 |
修稿时间: | 2003-07-16 |
Focal Length Measurement of Infrared Lenses |
| |
Affiliation: | (School of Electronic Engineering and Optic-electronic Technology,Nanjing University of Science and Technology,Nanjing 210094) |
| |
Abstract: | The image of a single slot diffraction is obtained using the pyroelectric vidicon(PEV) with the CO_2 laser as light source. According to the theory of single slot diffraction,the focal lengths of a set of the infrared Germanium lenses are measured. The result is presented. Some questions are discussed.The focal plane of the infrared lens is accurately decided by calculating the modutation transfer function (MTF). The method to validate the size of image capture system is discussed. |
| |
Keywords: | Infrared Gemanium lenses Focal length |