首页 | 本学科首页   官方微博 | 高级检索  
     


Comparison of Different Sorbents for Solid-Phase Extraction of Phenoxyalkanoic Acid Herbicides
Authors:Magdalena Biesaga  Anna Jankowska  Krystyna Pyrzyńska
Affiliation:(1) Department of Chemistry, University of Warsaw, Pasteura 1, PL-02-093 Warsaw, Poland
Abstract:An HPLC procedure for determination of phenoxyalkanoic acid herbicides in water samples is proposed. The analytical column Phenomenex C18(2) Luna 5 µm and UV detection at 225 nm were applied. Baseline resolution was achieved in isocratic mode with a mobile phase consisting of acetonitrile/acetic acid (40/60, v/v), adjusted to pH 2.5. SPE sorbents – C18 BondElut, phenyl-silica, LiChrolut SAX and polymeric sorbents – were compared for isolation and preconcentration of 6 phenoxyalkanoic acid herbicides. Higher (above 95%) and more reproducible recoveries were obtained with polymeric and phenyl-silica sorbents using pure methanol for elution. The method was tested for river water samples with the limit of detection in the range of 2–3 µg L−1 (for 50 mL sample) and a reproducibility of 5% RSD.
Keywords:: Chlorophenoxy acid herbicides   solid-phase extraction   water analysis.
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号