首页 | 本学科首页   官方微博 | 高级检索  
     


Fast HPLC method for the determination of glimepiride, glibenclamide, and related substances using monolithic column and flow program
Authors:El Deeb Sami  Schepers Udo  Wätzig Hermann
Affiliation:Institute of Pharmaceutical Chemistry, TU Braunschweig, Braunschweig, Germany.
Abstract:This work presents a fast method for the simultaneous separation and determination of glimepiride, glibenclamide, and two related substances by RP LC. The separation was performed on a Chromolith Performance (RP-18e, 100 mm x 4.6 mm) column. As mobile phase, a mixture of phosphate buffer pH 3, 7.4 mM, and ACN (55:45 v/v) was used. Column oven temperature was set to 30 degrees C. The total chromatographic run time was 80 s. This was achieved using a flow program from 5 to 9.9 mL/min. Precisions of the interday and the intraday assay for both retention times and peak areas for the four analyzed compounds were less than 1.2%. The method showed good linearity and recovery. The short analysis time makes the method very valuable for quality control and stability testing of drugs and their pharmaceutical preparations.
Keywords:Fast HPLC  Flow programming  Glibenclamide  Glimepiride  Monolithic columns