首页 | 本学科首页   官方微博 | 高级检索  
     检索      


PVC-based dicyclohexano-18-crown-6 sensor for La(III) ions
Authors:Mittal Susheel K  Kumar S K Ashok  Sharma Harish Kumar
Institution:School of Chemistry and Biochemistry, Thapar Institute of Engineering and Technology (Deemed University), Patiala 147004, India
Abstract:A new ion-selective electrode (ISE) based on dicyclohexano-18-crown-6 (DC18C6) as a neutral carrier is developed for lanthanum(III) ions. The electrode comprises of dicyclohexano-18-crown-6 (6%), PVC (33%), and ortho-nitrophenyl octyl ether (o-NPOE) (61%). The electrode shows a linear dynamic response in the concentration range of 10−6 to 10−1 M with a Nernstian slope of 19 mV per decade and a detection limit as 5×10−7 M. It has a response time of <30 s and can be used for at least 5 months without any significant divergence in potentials. The selectivity coefficients for mono-, di-, and trivalent cations indicate good selectivity for La(III) ions over a large number of interfering cations. The sensor has been used as an indicator electrode in the potentiometric titrations of La(III) with EDTA. The membrane is successfully applied in partially non-aqueous medium. It can be used in the pH range 4-9.
Keywords:DC18C6  Lanthanum(III) ions  Membrane electrode  Selectivity coefficients  Fixed interference method  Ion-selective electrode
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏