Determination of carbon dioxide in gaseous samples by gas diffusion-flow injection
Authors:Satienperakul Sakchai  Cardwell Terence J  Cattrall Robert W  McKelvie Ian D  Taylor Douglas M  Kolev Spas D
Institution:a Analytical Chemistry Laboratories, La Trobe University, Victoria 3086, Australia
b Water Studies Centre, School of Chemistry, Monash University, Victoria 3800, Australia
c School of Chemistry, The University of Melbourne, Victoria 3010, Australia
Abstract:A gas diffusion-flow injection system was developed for the determination of carbon dioxide in gaseous samples. The calibration was based on the use of either gaseous carbon dioxide or aqueous sodium carbonate standards. Gaseous carbon dioxide samples and gaseous or aqueous standards were injected directly into a donor stream of 1.0×10−4 M H2SO4. In the gas diffusion unit, carbon dioxide diffused through a PTFE membrane into an acceptor stream containing a mixed acid/base indicator. The absorbance of the acceptor stream was monitored spectrophotometrically at 554 nm. The calibration plot was linear over the range of 5.00×102 to 1.27×104 μl l−1 with a sample throughput of 28 h−1 and 3.2% R.S.D. (CO2]=2.37×103 μl l−1, n=12). The detection limit was determined as 2.50×102 μl l−1. The flow system was successfully applied to the analysis of several natural gaseous samples and the headspace of milk containers during storage. The flow injection results were found to be statistically indistinguishable at the 95% confidence level from those obtained by gas chromatography using thermal conductivity detection.
Keywords:Gas diffusion  Flow injection  Carbon dioxide determination  Gaseous samples  Bacterial spoilage  Soil atmosphere
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号