首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Use of the internal standardization for difficult sampling by graphite furnace atomic absorption spectrometry
Authors:Oliveira Adriana P  Neto José A Gomes  Nóbrega Joaquim A  Oliveira Pedro V
Institution:Departamento de Quimica Analitica, Universidade Estadual Paulista, P.O. Box 355, Araraquara-SP 14801-970, Brazil.
Abstract:This work shows the potentiality of As as internal standard to compensate errors from sampling of sparkling drinking water samples in the determination of selenium by graphite furnace atomic absorption spectrometry. The mixture Pd(NO3)2/Mg(NO3)2 was used as chemical modifier. All samples and reference solutions were automatically spiked with 500 μg l−1 As and 0.2% (v/v) HNO3 by the autosampler, eliminating the need for manual dilutions. For 10 μl dispensed sample into the graphite tube, a good correlation (r=0.9996) was obtained between the ratio of analyte absorbance by the internal standard absorbance and the analyte concentrations. The relative standard deviations (R.S.D.) of measurements varied from 0.05 to 2% and from 1.9 to 5% (n=12) with and without internal standardization, respectively. The limit of detection (LD) based on integrated absorbance was 3.0 μg l−1 Se. Recoveries in the 94-109% range for Se spiked samples were obtained. Internal standardization (IS) improved the repeatability of measurements and increased the lifetime of the graphite tube in ca. 15%.
Keywords:Internal standardization  Graphite furnace atomic absorption spectrometry  Complex matrices  Sparkling waters
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏