Use of high-voltage electroaccumulation for lowering the detection limit for silver in stripping voltammetry |
| |
Authors: | L I Skvortsova V N Kiryushov |
| |
Institution: | (1) Institute of Solid-State Chemistry and Mechanochemistry, Siberian Division, Russian Academy of Sciences, ul. Kutateladze 18, 630128 Novosibirsk, Russia |
| |
Abstract: | The use of high-voltage electroaccumulation in the determination of silver by stripping voltammetry at a mechanically renewed
graphite electrode with an area of 2.5 × 10−3 cm2 in 0.2% boric acid acidified with acetic acid at a voltage of 400 V and at an accumulation time of 5 min lowered the detection
limit for silver to 1 × 10−12 M. |
| |
Keywords: | |
本文献已被 SpringerLink 等数据库收录! |