首页 | 本学科首页   官方微博 | 高级检索  
     


A microgap impedance sensor for the determination of trace water in organic solvents
Authors:Liang Jie  Chen Bao-Qin  Long Yi-Tao
Affiliation:Shanghai Key Laboratory of Functional Materials Chemistry & Department of Chemistry, East China University of Science and Technology, Shanghai, 200237, China.
Abstract:A microgap impedance sensor with a 50 μm gap was developed for the determination of trace water in organic solvents by coating poly(dimethyldiallylammonium chloride) (PDMDAAC) and ferricyanide/ferrocyanide composite materials on indium tin oxide (ITO). The electrochemical properties of the composite materials were investigated with cyclic voltammetry (CV) and electrochemical impedance spectroscopy (EIS). We observed that the impedance response of the sensor depended on the concentration of trace water in the organic solvents. Under optimized conditions, the linear range for the determination of trace water was 0-0.06% for chloroform (CHCl(3)), 0-0.10% for acetone (CH(3)COCH(3)), 0-0.12% for tetrahydrofuran (THF), and 0-0.10% for acetonitrile (CH(3)CN), and the detection limits were 0.65, 1.54, 0.61, and 1.72 ppm, respectively. The results obtained from the impedance sensors were comparable to those obtained using the traditional Karl Fischer method.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏