Takasaki Radiation Chemistry Research Establishment, Advanced Radiation Technology Centre, Japan Atomic Energy Research Institute, Environmental Conservation Process Laboratory, 1233 Watanuki-machi, Takashi-shi, Gumna 370-1292, Japan
Abstract:
Four kinds of film dosimeters well-characterized for low linear energy transfer (LET) radiations were applied to 3–45 MeV/u ions. The dose–responses relative to those for low LET radiations are almost one at a stopping power of about 10 MeV/(mg/cm2) and gradually become smaller with an increase in the stopping power. The overall uncertainty in ion beam dosimetry using these characterized dosimeters is better than ±5% (1σ) including uncertainty in fluence measurement (±2%). Lateral- and depth-dose profile measurements were made using one of the dosimeters (Gafchromic) with a spatial resolution of better than 1 and 10 μm, respectively.