首页 | 本学科首页   官方微博 | 高级检索  
     


3–45 MeV/u ion beam dosimetry using thin film dosimeters
Authors:T. Kojima   H. Sunaga   H. Takizawa   H. Hanaya  H. Tachibana
Affiliation:

Takasaki Radiation Chemistry Research Establishment, Advanced Radiation Technology Centre, Japan Atomic Energy Research Institute, Environmental Conservation Process Laboratory, 1233 Watanuki-machi, Takashi-shi, Gumna 370-1292, Japan

Abstract:Four kinds of film dosimeters well-characterized for low linear energy transfer (LET) radiations were applied to 3–45 MeV/u ions. The dose–responses relative to those for low LET radiations are almost one at a stopping power of about 10 MeV/(mg/cm2) and gradually become smaller with an increase in the stopping power. The overall uncertainty in ion beam dosimetry using these characterized dosimeters is better than ±5% (1σ) including uncertainty in fluence measurement (±2%). Lateral- and depth-dose profile measurements were made using one of the dosimeters (Gafchromic) with a spatial resolution of better than 1 and 10 μm, respectively.
Keywords:Thin film dosimeters   Relative effectiveness   Ion beams   Mass collision stopping power   Uncertainty   Depth-dose profile   Lateral dose profile   Spatial resolution
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏