Origin and correction of the deviations in retention times at increasing flow rate with Chromolith columns
Authors:S. Pous-Torres,J.R. Torres-Lapasió  M.J. Ruiz-Á  ngel,M.C. Garcí  a-Á  lvarez-Coque
Affiliation:Departament de Química Analítica, Universitat de València, c/Dr. Moliner 50, 46100 Burjassot, Spain
Abstract:Chromoliths can be used at flow rates beyond those feasible for conventional microparticulate packed columns. Ideally, the plots of the retention time versus the inverse of delivered flow rate should exhibit y-intercept of zero. However, significant positive deviations correlating with the solute polarity were observed for several compounds chromatographed with a Chromolith column, owing to the increased system pressure. Consequently, the dead time marker exhibits a smaller deviation, making the retention factors depend on the flow rate. Chromoliths are made of a silica-based monolith encapsulated within a PEEK tube, and should suffer larger stress with pressure than stainless steel columns, tending to inflate them and increase their volume. This decreases the linear velocity inside the column, and increases the retention at relatively low pressure (<200 bar). In contrast, frictional heating, which is an issue for microparticulate columns, seems to be less significant for the highly permeable Chromoliths. The usefulness of the retention time versus the inverse of the delivered flow rate plots to measure the deviations, whatever their origin, is shown. This allows the correction of the retention times to the ideal behaviour, where the retention factors are independent of the flow rate.
Keywords:HPLC   Chromolith   XTerra   Flow rate   Pressure effects   Retention time deviations
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号