首页 | 本学科首页   官方微博 | 高级检索  
     检索      


HPLC Determination and Pharmacokinetic Study of Valdecoxib in Human Plasma
Authors:R T Sane  S Menon  A Y Deshpande  A Jain
Institution:(1) Department of Chemistry, Ramnarain Ruia College, Matunga, Mumbai, 400019, India;(2) Therapeutic Drug Monitoring Laboratory, 194, Scheme No. 6, Road No.15, Sion (E), Mumbai, 400 022, India
Abstract:A sensitive, simple, and accurate high-performance liquid chromatographic method has been developed for determination of valdecoxib and the internal standard rofecoxib in human plasma. Protein was precipitated from plasma samples by addition of perchloric acid (HClO4); the drug was then extracted with diethyl ether. Separation was performed on a Cosmosil C18 column (150 mm × 4.6 mm i.d., 5 mgrm particles) with ammonium acetate buffer-acetonitrile, 60:40 (v/v), containing 0.1% TEA, pH 6.5, as mobile phase. Detection and quantification were performed by UV-visible detection at 239 nm. Detection and quantification limits were 3 and 5 ng mL–1, respectively. The linear concentration range for valdecoxib was 5–400 ng mL–1. The validated RP HPLC method was used for determination of the pharmacokinetic data for the drug in humans.
Keywords:Column liquid chromatography  Valdecoxib  Rofecoxib  Human plasma
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏