首页 | 本学科首页   官方微博 | 高级检索  
     


SI-traceable values for cadmium concentration in the water samples of IMEP-6
Authors:I. Papadakis  P. D. P. Taylor  P. D. Bièvre
Affiliation:(1) Institute for Reference Materials and Measurements, European Commission-JRC, B-2440 Geel, Belgium, BE
Abstract: The certification measurement for the cadmium concentration in the samples as performed by Isotope Dilution (ID) using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) is described in framework of the International Measurement Evaluation Programme (IMEP). By using a primary method of measurement (ID), making up a full uncertainty budget and using a sensitive technique (ICP-MS), reference values traceable to SI can be obtained for this material. The uncertainty budget is performed in accordance with the ISO and EURACHEM guides on uncertainty. A comparison is made with certification results of other experienced laboratories also using primary methods where possible. Received: 8 July 1996/Revised 26 August 1996/Accepted: 1 September 1996
Keywords:
本文献已被 SpringerLink 等数据库收录!