Compact fluorescence detection using in-fiber microchannels-its potential for lab-on-a-chip applications |
| |
Authors: | Irawan Rudi Tay Chia Meng Tjin Swee Chuan Fu Chit Yaw |
| |
Affiliation: | Singapore-University of Washington Alliance, Biomedical Engineering Research Centre, Nanyang Technological University, Singapore. erirawan@ntu.edu.sg |
| |
Abstract: | This paper reports a compact and practical fluorescence sensor using an in-fiber microchannel. A blue LED, a multimode PMMA or silica fiber and a mini-PMT were used as an excitation source, a light guide and a fluorescence detector, respectively. Microfluidic channels of 100 microm width and 210 microm depth were fabricated in the optical fibers using a direct-write CO(2) laser system. The experimental results show that the sensor has high sensitivity, able to detect 0.005 microg L(-1) of fluorescein in the PBS solution, and the results are reproducible. The results also show that the silica fiber sensor has better sensitivity than that of the PMMA fiber sensor. This could be due to the fouling effect of the frosty layer formed at the microchannel made within the PMMA fiber. It is believed that this fiber sensor has the potential to be integrated into microfluidic chips for lab-on-a-chip applications. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|