A simplified methodology: pH sensing using an in situ fabricated Ir electrode under neutral conditions
Authors:Chen  Yuqi  Li  Xiuting  Li  Danlei  Batchelor-McAuley  Christopher  Compton  Richard G
Institution:1.Institute for Advanced Study, Shenzhen University, Shenzhen, 518060, China
;2.Department of Chemistry, Physical and Theoretical Chemistry Laboratory, University of Oxford, South Parks Road, Oxford, OX1 3QZ, UK
;
Abstract:

Herein, a simplified fabrication method for the producing of a pH-sensitive iridium electrode is developed. The in situ electrochemical fabrication of an iridium oxide film is optimized and shown to be achievable under neutral conditions rather than the acidic conditions hitherto employed. The formation of a pH sensitive Ir(III/IV) hydrous film is confirmed via XPS. The amperometric pH-sensing properties of this electrochemically generated material were investigated using square wave voltammetry. In the pH range 2–13, the iridium oxide redox signal has a pH dependency of 86.1 ± 1.1 mV per pH unit for midpoint potentials with uncertainties being ± 0.01–0.05 pH. Finally, the newly developed pH sensor was used to measure the pH of a natural water sample with excellent results as compared to a conventional glass pH probe.

Keywords: