首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Error analysis for a Mach-Zehnder type speckle interferometer
Authors:Pascal Picart
Institution:

a École Nationale Supérieure d’Ingénieurs du Mans, rue Aristote, 72085 Le Mans Cedex 9, France

b Laboratoire d’Acoustique de l’Université du Maine, UMR CNRS 6613, Avenue Olivier Messiaen, 72085 Le Mans Cedex 9, France

Abstract:A phase shifting speckle interferometer based on a Mach-Zehnder geometry is presented. The displacement field is measured using an electro-optic phase shifting technique. An analytical investigation of the accuracy of the interferometer is performed by considering the most contributing error sources. It is shown that in the case of the use of a low power laser source, the geometrical aberration of the optical component is the main contribution to the systematic errors. The case of the sampling of a full measurement is investigated. It is demonstrated that the systematic error of the full measurement has the same statistics as those given by a simple acquisition. Experimental results are reported in the case of the measurement of the deformation of industrial connectors submitted to a crushing test. It is found that for a peak to valley of 32 μm, the trueness of the measurement appears higher than 160 nm.
Keywords:Speckle interferometry  Phase shifting  Mach-Zehnder interferometer  Polarisation  Systematic errors  Geometrical aberrations
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏